Global Sites:  Australia (English)

Skip navigation links
    

Solutions that drive knowledge.

 

 

 

Solutions that drive knowledge.

Clinical Laboratories | VITROS® 5,1 FS

 



 

Specifications

Reagents: No preparation required *; on-board stability up to 35 days and shelf life stability up to 18 months from date of manufacture

MicroSensorTM: Scan for sample interferences (hemolysis, icterus, turbidity) over 400-800 nm wavelengths 

  No additional reagents, sample or consumables required
  Analysis time is less than 1 second
  Automatic flagging of impacted tests  

On-Board Test Capacity:

  Up to 80 different assays resident onboard simultaneously
  89 MicroSlideTM and 36 MicroTipTM positions

Measurement Principles: Potentiometric (direct ISE), Colorimetric/Rate, Immuno-rate, Turbidimetric  

  MicroSlide wavelengths: 340, 400, 460, 540, 600, 630, 670 & 680 nm
  MicroTip wavelengths: 340, 380, 405, 510, 540, 575, 600, 620, 660, 700 & 800 nm

Calibration: Unlimited number of lots may be pre-calibrated
                    
Stability: Up to 6 months 
                    Process: Random access;25 calibrations per assay body fluid  
                    combination 
                    Types: linear, Logit 4, Logit 5 and cubic spline algorithms
Instrument Startup: No priming; no washes; no daily calibration or calibration checks

Time for Single Result: 

Potentiometric: ~2.5 minutes
Colorimetric: ~7 minutes
Immuno-Rate: ~9 minutes
MicroTip: ~8-16 minutes

                        Real-time results data stream to LIS
Sample Types: Serum, Plasma, Urine, CSF, Whole Blood **
Sample Volume: 2-12 µL; Dead volume: minimum 35 µL
Sample Capacity: 160 samples in Universal Sample Trays (150 routine samples plus 10 in dedicated STAT lane)

Sample & Reagent Management: Intellicheck™ Technology includes clot and bubble detection, Save-the-Sample Clot Management, liquid level sensing and reagent delivery detection and management

Universal Sample Tray: Holds samples in 1.5 mL, 5 mL, 7 mL and 10 mL collection tubes, 0.1 mL Microtainer , Microsample cup and 0.5 & 2 mL cups

Barcode Types: Autodiscriminates among all standard symbologies, including Code 128, Code 39, Codabar, Interleaved 2 of 5

Barcode Placement: 3D barcode reader allows deviation by +/- 15 degrees from vertical

Operator Interface: Graphical user interface, Ergonomic Touchscreen Monitor with on-board documentation and animated HELP (V-DOCS); 32 programmable panels

User-Defined Channels: 10 available

Data Storage: 25,000 samples, ability to archive and retrieve through CD-ROM

Instrument Dimensions:
                       Length: 92.2 inches / 2.34 m
                       Width: 33.4 inches / 0.84 m
                       Height: 52.5 inches / 1.33 m
                       Weight: ~1,400 pounds / ~640 kg
  

2 Electrical

Environment

Communications


Line Voltage: Dedicated 20 amperes, nominal 200-240V

Line Frequency: 47-63 Hz


Average Heat Output: 4, 187 BTU

 

Operating Temperature Range: 59-86 degree F / 15-30 degrees C

 

Ambient Humidity: 15-75% RH

 

Altitude Limit: Up to 2,438 meters/ 8,000 feet

Plumbing: No water
or drain required


LIS Interface Capability: ASTM

 

LIS Download Process: Dynamic Download and Host Query

 

Ports: 1 RS 232 serial port; 2

USB ports; 2 Ethernet ports

 

e-Connectivity® Interactive System Management:

a Using DSL, cable modem or intranet through a VPN using QNX software
a Includes online secure remote diagnostics, bi-directional communication and software download capability

 

Laboratory Information System
Information about the VITROS®  5,1 FS LIS system can be found in our Technical Resource Library by clicking the "other technical documentation" link under the "VITROS®  Chemistry Systems" category. 
 

Electrical
Line Voltage: Dedicated 20 amperes, nominal 200-240V
Line Frequency: 47-63 Hz  

Environment
Average Heat Output: 4, 187 BTU
Operating Temperature Range: 59-86 degree F / 15-30 degrees C
Ambient Humidity: 15-75% RH
Altitude Limit: Up to 2,438 meters/ 8,000 feet
Plumbing: No water or drain required 

Communications
LIS Interface Capability: ASTM
LIS Download Process: Dynamic Download and Host Query
Ports: 1 RS 232 serial port; 2 USB ports; 2 Ethernet ports  

e-Connectivity® Interactive System Management
Using DSL, cable modem or intranet through a VPN using QNX software
Includes online secure remote diagnostics, bi-directional communication and software download capability

* Except Caffeine
** For Auto %Hemoglobin A1c
*** Available post launch

                       


                   
VITROS® 5,1 FS



     

This site is published by Ortho Clinical Diagnostics, which is part of Johnson & Johnson Medical Pty Limited which is solely responsible for its contents. © All contents copyright Ortho Clinical Diagnostics which is part of Johnson & Johnson Medical Pty Limited, 2003-2011 unless otherwise noted. All rights reserved. Product availability is subject to fulfillment of regulatory requirements in each market. This site was written for healthcare professionals in Australia only. See our Legal Notice and Privacy Policy.

Last Modified Date:18/07/2009 9:12